Usefulness of non-lead aprons in radiation protection for physicians performing interventional procedures.
Usefulness of non-lead aprons in radiation protection for physicians performing interventional procedures.
复制标题
DOI:
10.1093/rpd/ncn244
复制
发表时间:
2008-09
影响因子:
1
通讯作者:
M. Zuguchi;K. Chida;Masaaki Taura;Y. Inaba;Ayako Ebata;S. Yamada
中科院分区:
文献类型:
--
作者:
M. Zuguchi;K. Chida;Masaaki Taura;Y. Inaba;Ayako Ebata;S. Yamada
At present, interventional radiology (IVR) tends to involve long procedures (long radiation duration), and physicians are near to the source of scattered radiation. Hence, shielding is critical in protecting physicians from radiation. Protective aprons and additional lead-shielding devices, such as tableside lead drapes, are important means of protecting the physician from scattered radiation. The purpose of this study was to evaluate whether non-lead aprons are effective in protecting physicians from radiation during IVR procedures. In this study, the radiation protection effects of commercially available protective lead and non-lead aprons, when exposed to diagnostic X rays, are compared. The performance of these non-lead and lead aprons was similar for scattered X rays at tube voltages of 60-120 kV. Properly designed non-lead aprons are thus more suitable for physicians because they weigh approximately 20% less than the lead aprons, and are non-toxic.