Inappropriate Shocks Delivered by an ICD as a Result of Sensed Potentials from a Transcutaneous Electronic Nerve Stimulation Unit
Inappropriate Shocks Delivered by an ICD as a Result of Sensed Potentials from a Transcutaneous Electronic Nerve Stimulation Unit
复制标题
由于经皮电子神经刺激装置感应到电势而由 ICD 发出不适当的电击
DOI:
10.1111/j.1540-8159.1998.tb00027.x
复制
发表时间:
1998
期刊:
影响因子:
--
通讯作者:
M. Schoenfeld
中科院分区:
文献类型:
--
作者:
D. Philbin;M. Marieb;KESHAVA H. Aithal;M. Schoenfeld
Although the potential for transcutaneous electronic nerve stimulation (TENS) units to interfere with appropriate function of cardiac pacemakers is well documented, an adverse interaction between a TENS unit and an ICD has never been reported. We describe a patient in whom a TENS unit created an electrical artifact that was interpreted by a transvenous ICD as ventricular fibrillation, leading to the delivery of inappropriate therapy. TENS units should be used with caution in patients with ICDs.