Method for calibrating a Rogowski coil of fast time response.

Method for calibrating a Rogowski coil of fast time response.
复制标题

DOI:
10.1063/1.2764373
复制
发表时间:
2007-08
期刊:
The Review of scientific instruments
影响因子:
--
通讯作者:
Rui Liu;Xinxin Wang;X. Zou;Jianqiang Yuan;Naigong Zou;Luya He
Rui Liu;Xinxin Wang;X. Zou;Jianqiang Yuan;Naigong Zou;Luya He
中科院分区:
其他
文献类型:
--
作者:
Rui Liu;Xinxin Wang;X. Zou;Jianqiang Yuan;Naigong Zou;Luya He

文献摘要

被引文献

相似文献

设计了一种新的校准快时间响应Rogowski线圈的方法。该方法是基于电缆脉冲发生器的方法,除了电压信号传感器输出被移动到一个位置与距离l的负载。如果2l/v比正向电压脉冲u(f)(t)的持续时间长,则反射电压脉冲u(r)(t)可以与u(f)(t)分离并直接测量。利用公式i(t)=[u(f)(t)-u(r)(t)]/50计算Rogowski线圈的初级电流,可以更准确地标定线圈。
A new method was designed for calibrating a Rogowski coil of fast time response. The method is based on the cable pulser method except that the voltage signal pick-off output was moved to a position with a distance l from the load. If 2l/v is longer than the time duration of the forward voltage pulse u(f)(t), the reflected voltage pulse u(r)(t) could be separated from u(f)(t) and directly measured. Using the formula i(t)=[u(f)(t)-u(r)(t)]/50 to calculate the primary current of the Rogowski coil, the coil could be more accurately calibrated.