Gradient waveform pre-emphasis based on the gradient system transfer function
Gradient waveform pre-emphasis based on the gradient system transfer function
复制标题
DOI:
10.1002/mrm.27147
复制
发表时间:
2018-10-01
影响因子:
3.3
通讯作者:
Koestler, Herbert
中科院分区:
文献类型:
--
作者:
Stich, Manuel;Wech, Tobias;Koestler, Herbert
Purpose: The gradient system transfer function (GSTF) has been used to describe the distorted k-space trajectory for image reconstruction. The purpose of this work was to use the GSTF to determine the pre-emphasis for an undistorted gradient output and intended k-space trajectory.Methods: The GSTF of the MR system was determined using only standard MR hardware without special equipment such as field probes or a field camera. The GSTF was used for trajectory prediction in image reconstruction and for a gradient waveform pre-emphasis. As test sequences, a gradient-echo sequence with phase-encoding gradient modulation and a gradient-echo sequence with a spiral read-out trajectory were implemented and subsequently applied on a structural phantom and in vivo head measurements.Results: Image artifacts were successfully suppressed by applying the GSTF-based pre-emphasis. Equivalent results are achieved with images acquired using GSTF-based post-correction of the trajectory as a part of image reconstruction. In contrast, the pre-emphasis approach allows reconstruction using the initially intended trajectory.Conclusion: The artifact suppression shown for two sequences demonstrates that the GSTF can serve for a novel pre-emphasis. A pre-emphasis based on the GSTF information can be applied to any arbitrary sequence type.