MAGNETIC-FIELD MAPPING
MAGNETIC-FIELD MAPPING
复制标题
DOI:
10.1002/mrm.1910060204
复制
发表时间:
1988-02-01
影响因子:
3.3
通讯作者:
BRYANT, RG
中科院分区:
文献类型:
--
作者:
HORNAK, JP;SZUMOWSKI, J;BRYANT, RG
Homogeneous radiofrequency magnetic fileds are necessary for production of high-quality magnetic images and for most forms of magnetic resonance spectroscopy. It is often convenient to map the radiofrequency homogeneity associated with a resonant device by measuring the magnetic image intensity of an aqueous phantom placed within the resonator. The rf field intensity is not related trivially to the magnetic image intensity, and the relationship is different for different image acquisition methods. In this report relationships between rf field intensity and magnetic image intensity are derived and radiofrequency field maps presented for comparison using an rf probe moved about within the resonant volume, spin-echo images, and small tip angle gradient refocused echo images.