Transcutaneous measurement of renal function in conscious mice
Transcutaneous measurement of renal function in conscious mice
复制标题
DOI:
10.1152/ajprenal.00279.2012
复制
发表时间:
2012-09-01
影响因子:
4.2
通讯作者:
Schock-Kusch, Daniel
中科院分区:
文献类型:
--
作者:
Schreiber, Andrea;Shulhevich, Yury;Schock-Kusch, Daniel
Schreiber A, Shulhevich Y, Geraci S, Hesser J, Stsepankou D, Neudecker S, Koenig S, Heinrich R, Hoecklin F, Pill J, Friedemann J, Schweda F, Gretz N, Schock-Kusch D. Transcutaneous measurement of renal function in conscious mice. Am J Physiol Renal Physiol 303: F783-F788, 2012. First published June 13, 2012; doi: 10.1152/ajprenal.00279.2012.-Determination of glomerular filtration rate (GFR) in conscious mice is cumbersome for the experimenter and stressful for the animals. Here we report on a simple new technique allowing the transcutaneous measurement of GFR in conscious mice. This approach extends our previously developed technique for rats to mice. The technique relies on a miniaturized device equipped with an internal memory that permits the transcutaneous measurement of the elimination kinetics of the fluorescent renal marker FITC-sinistrin. This device is described and validated compared with FITC-sinistrin plasma clearance in healthy, unilaterally nephrectomized and pcy mice. In summary, we describe a technique allowing the measurement of renal function in freely moving mice independent of blood or urine sampling as well as of laboratory assays.