Doppler velocity detection limitations in spectrometer-based versus swept-source optical coherence tomography.

Doppler velocity detection limitations in spectrometer-based versus swept-source optical coherence tomography.
复制标题

DOI:
10.1364/boe.2.002175
复制
发表时间:
2011-08-01
影响因子:
3.4
通讯作者:
Izatt JA
Izatt JA
中科院分区:
医学2区
文献类型:
--
作者:
Hendargo HC;McNabb RP;Dhalla AH;Shepherd N;Izatt JA

文献摘要

被引文献

相似文献

Recent advances in Doppler techniques have enabled high sensitivity imaging of biological flow to measure blood velocities and vascular perfusion. Here we compare spectrometer-based and wavelength-swept Doppler OCT implementations theoretically and experimentally, characterizing the lower and upper observable velocity limits in each configuration. We specifically characterize the washout limit for Doppler OCT, the velocity at which signal degradation results in loss of flow information, which is valid for both quantitative and qualitative flow imaging techniques. We also clearly differentiate the washout effect from the separate phenomenon of phase wrapping. We demonstrate that the maximum detectable Doppler velocity is determined by the fringe washout limit and not phase wrapping. Both theory and experimental results from phantom flow data and retinal blood flow data demonstrate the superiority of the swept-source technique for imaging vessels with high flow rates.