A REVIEW OF TWO METHODS USED IN THE USA TO ASSESS HE DURING FLUOROSCOPIC-BASED RADIOLOGY
A REVIEW OF TWO METHODS USED IN THE USA TO ASSESS HE DURING FLUOROSCOPIC-BASED RADIOLOGY
复制标题
DOI:
10.1093/rpd/ncw006
复制
发表时间:
2016-09-01
影响因子:
1
通讯作者:
Salasky, Mark R.
中科院分区:
文献类型:
--
作者:
Yoder, R. Craig;Salasky, Mark R.
Dosemeter results for similar to 81 500 people performing fluoroscopic and interventional radiology procedures were examined to identify differences between groups monitored either by using two dosemeters, one placed at the collar above the apron and a second placed under the apron on the torso (EDE1) or by using one single dosemeter placed at the collar above the apron (EDE2). The median annual H-E was 0.17 mSv for those monitored using the EDE1 protocol and 0.26 mSv for the group using the EDE2 protocol. The EDE2 method was used most frequently with the EDE1 method preferred for those more highly exposed. Approximately, 22 % of dosemeter results for EDE1 were inconsistent with expected norms based on over and under apron dosemeter relationships.