A Gel Pad Designed to Measure Muscle Volume Using Freehand 3‐Dimensional Ultrasonography
A Gel Pad Designed to Measure Muscle Volume Using Freehand 3‐Dimensional Ultrasonography
复制标题
设计用于使用徒手 3 维超声检查测量肌肉体积的凝胶垫
DOI:
10.1002/jum.15490
复制
发表时间:
2020
影响因子:
2.3
通讯作者:
A. Nordez
中科院分区:
文献类型:
--
作者:
Julien Thomare;L. Lacourpaille;P. McNair;Marion Crouzier;R. Ellis;A. Nordez
We developed an innovative gel pad that covers the entire lower leg to remove artifacts due to the pressure of the transducer in freehand 3‐dimensional ultrasonography. In comparison to the reference method in water, this study showed that this new method was valid (bias, 3.4 mL; limit of agreement, 7.7 mL for a volume of ≈220 mL) and reliable (coefficient of variation, <1.1%) for the measurement of gastrocnemius medialis muscle volume. Considering that it is easier to use than the water tank technique, it has much promise for volumetric measurement of many muscles.