An RPC-PET prototype with high spatial resolution
An RPC-PET prototype with high spatial resolution
复制标题
DOI:
10.1016/j.nima.2004.07.016
复制
发表时间:
2004-11
影响因子:
1.4
通讯作者:
A. Blanco;N. Carolino;C. Correia;R. Marques;P. Fonte;D. González-Díaz;A. Lindote;M. Lopes;M. P. Macedo;A. Policarpo
中科院分区:
文献类型:
--
作者:
A. Blanco;N. Carolino;C. Correia;R. Marques;P. Fonte;D. González-Díaz;A. Lindote;M. Lopes;M. P. Macedo;A. Policarpo
A small positron emission tomography system, based on the timing RPC technology has been built and tested. This first prototype is aimed at validating the expectations, derived from simulations, of a very high spatial resolution, which could be of value for the imaging of small animals. The system is composed of two counting heads, able to measure the photon interaction point in two dimensions, the transaxial dimension and the Depth of Interaction. Each head is composed of 16 independent stacked RPCs made from copper and glass (anode) electrodes. Point-like22Na sources were so far reconstructed, using the standard filtered back-projection algorithm, with a spatial accuracy of 0.6mm FWHM, free of parallax error.