Study of the effect of the surface state on the cytocompatibility of a Co-Cr alloy using human osteoblasts and fibroblasts.
Study of the effect of the surface state on the cytocompatibility of a Co-Cr alloy using human osteoblasts and fibroblasts.
复制标题
使用人成骨细胞和成纤维细胞研究表面状态对钴铬合金细胞相容性的影响。
DOI:
--
复制
发表时间:
1990
期刊:
影响因子:
--
通讯作者:
M. Harmand
中科院分区:
文献类型:
--
作者:
A. Naji;M. Harmand
Cobalt-chromium-based alloys are widely used in oral and orthopedic implantology. Although they are relatively well tolerated, biological complications could occur which sometimes are due to the insufficient biocompatibility of the alloy. This study shows the effects of an alloy (Co (base), 28% Cr, 5.5% Mo, 1% Ni, 0.95% Si, 0.7% Fe, 0.65% Mn, 0.25% C), on differentiated human cells derived from an oral implantation site, specifically alveolar bone osteoblasts and gingival fibroblasts. The cytocompatibility of the alloy is determined by the study of cell proliferation, determination of total cell protein and intracellular alkaline phosphatase contents, cytoskeleton, and cell morphology. The alloy is presented to the cells in four different surface states: rough cast, specular polished, microbead blasted, and RF sputtered. The results demonstrate that the same material has different effects on the basal and specific cellular functions, according to its surface state. For this alloy we can classify its cytocompatibility according to its surface state in such an order: Microbead blasted much greater than specular polished greater than RF sputtered greater than rough cast.
影响因子:
14
作者:
Mueller,HJ
通讯作者:
Mueller,HJ