Radiation tolerance of CVD diamond detectors for pions and protons
Radiation tolerance of CVD diamond detectors for pions and protons
复制标题
介子和质子 CVD 金刚石探测器的辐射耐受性
DOI:
10.1016/s0168-9002(01)01666-7
复制
发表时间:
2002
影响因子:
1.4
通讯作者:
M. Zoeller
中科院分区:
文献类型:
--
作者:
W. Adam;E. Berdermann;P. Bergonzo;G. Bertuccio;F. Bogani;E. Borchi;A. Brambilla;M. Bruzzi;C. Colledani;J. Conway;P. D’Angelo;W. Dąbrowski;P. Delpierre;A. Deneuville;W. Duliński;B. Eijk;A. Fallou;F. Fizzotti;F. Foulon;M. Friedl;K. Gan;E. Gheeraert;G. Hallewell;S. Han;F. Hartjes;J. Hrubec;D. Husson;H. Kagan;D. Kania;J. Kapłon;R. Kass;T. Koeth;M. Krammer;A. Logiudice;R. Lu;L. Lynne;C. Manfredotti;D. Meier;M. Mishina;L. Moroni;J. Noomen;A. Oh;L. Pan;M. Pernicka;A. Peitz;L. Perera;S. Pirollo;M. Procario;J. Riester;S. Roe;L. Rousseau;A. Rudge;J. Russ;S. Sala;M. Sampietro;S. Schnetzer;S. Sciortino;H. Stelzer;R. Stone;B. Suter;R. Tapper;R. Tesarek;W. Trischuk;D. Tromson;E. Vittone;A. Walsh;R. Wedenig;P. Weilhammer;M. Wetstein;C. White;W. Zeuner;M. Zoeller
The paper gives new results on the radiation tolerance of CVD diamond for irradiation with 300MeV/c pions and 24GeV/c protons. The measured charge signal spectrum is compared at several irradiation levels with the spectrum calculated by a model. Irradiation by particles causes damage leading to a decrease of the charge signal. However, both the measurements and the outcome from the model show that for tracker applications this drawback is at least partly counterbalanced by a narrowing of the distribution curve of the charge signal. As a result, the efficiency of a CVD diamond tracker is less affected by irradiation than the mean charge signal.