Performance of Vienna Type Differential Mobility Analyzer at 1.2-20 Nanometer
Performance of Vienna Type Differential Mobility Analyzer at 1.2-20 Nanometer
复制标题
1.2-20纳米Vienna型差分淌度分析仪的性能
DOI:
10.1080/02786829708965503
复制
发表时间:
1997
影响因子:
5.2
通讯作者:
J. Necid
中科院分区:
文献类型:
--
作者:
G. Reischl;J. Mäkelä;J. Necid
ABSTRACT The transfer functions of the so called “Vienna type” differential mobility analyzer (DMA) were studied experimentally at the size range of 1.2–20 nm following the procedure suggested by Stolzenburg (1988) Ph.D. Thesis. For the study, two identical DMAs were used as a tandem DMA (TDMA) arrangement to measure nanoparticles from a hot wire WOx generator. The approximate analytical formulae by Stolzenburg describing diffusion broadening of the DMA transfer function were observed to be valid with high accuracy from 20 nm down to ∼ 2 nm. In the size range of 1.2–2 nm nominal mobility equivalent diameter small deviations of the shape of the TDMA output signal predicted by Stolzenburg were found.