A one step vs. a multi step geometric calibration of an optical coherence tomography
A one step vs. a multi step geometric calibration of an optical coherence tomography
复制标题
光学相干断层扫描的一步与多步几何校准
DOI:
10.1117/12.2005264
复制
发表时间:
2013
影响因子:
4.8
通讯作者:
T. Ortmaier
中科院分区:
文献类型:
--
作者:
J. Díaz Díaz;M. Rahlves;O. Majdani;E. Reithmeier;T. Ortmaier
This contribution compares two approaches for the geometric calibration of an optical coherence tomography (OCT) which forms part of a medical navigation system. For this purpose, a one step and a multi step calibration is performed with a self-produced 3D reference structure and a high-accurate 6 degrees of freedom (DoF) parallel robot, respectively. These 3D landmark-based geometric calibrations are based on the identification of a parameterized grey-box OCT model. We show in experimental results that both methods reduce systematic errors by more than one order of magnitude.