X-ray micro-CT with a displaced detector array

X-ray micro-CT with a displaced detector array
复制标题

DOI:
10.1118/1.1489043
复制
发表时间:
2002-07-01
期刊:
影响因子:
3.8
通讯作者:
Wang, G
Wang, G
中科院分区:
医学3区
文献类型:
--
作者:
Wang, G

文献摘要

被引文献

相似文献

由于x射线微ct应用中样品的大小不同,因此希望有一种机制来改变微ct扫描仪的视野。将视场直径扩大一倍的一种众所周知的方法是将探测器阵列移动50%。在本文中,我们提出了探测器阵列位移大于0%但小于50%的连续可调视场,并制定了一种无伪影重建的加权方案。然后,我们用Shepp-Logan模型进行了数值模拟,以证明在扇形梁和锥形梁几何结构下的可行性。(C) 2002年美国医学物理学家协会。
Because the sizes of samples differ in x-ray micro-CT applications, it is desirable to have a mechanism to change the field of view of a micro-CT scanner. A well-known way to double the diameter of the field of view is to displace a detector array by 50%. In this paper, we propose to displace a detector array by an amount of greater than 0% but less than 50% for a continuously adjustable field of view, and formulate a weighting scheme for artifact-free reconstruction. Then, we perform numerical simulation with the Shepp-Logan phantom to demonstrate the feasibility in fan-beam and cone-beam geometry. (C) 2002 American Association of Physicists in Medicine.