STAN in clinical practice -: The outcome of 2 years of regular use in the city of Gothenburg
STAN in clinical practice -: The outcome of 2 years of regular use in the city of Gothenburg
复制标题
DOI:
10.1016/j.ajog.2006.01.108
复制
发表时间:
2006-07-01
影响因子:
9.8
通讯作者:
Hagberg, Henrik
中科院分区:
文献类型:
--
作者:
Noren, Hakan;Blad, Sofia;Hagberg, Henrik
Objective: The purpose of this study was to monitor the introduction of the STAN-methodology (Noventa Medical, Moelndal, Sweden).Study design: This was a prospective observational study covering the total population of deliveries at term during 2 years. Four thousand eight hundred and thirty out of 14,687 term pregnancies were monitored using the STAN S 21 fetal heart monitor and the associated clinical guidelines. Cord artery metabolic acidosis, neonatal outcome, and rates of operative deliveries for fetal distress were assessed.Results: The annual rate of STAN usage increased from 28.1% to 37.7% and was associated with a significant reduction in metabolic acidosis rate in the total population from 0.76% to 0.44% (P < .05). The compliance with the clinical guidelines increased in cases requiring intervention. The rates for moderate/severe hypoxic neonatal encephalopathy were consistently low, 0.55 and 0.68 per 1000 deliveries, respectively, and corresponding to previous findings. The rate of operative delivery did not change during the 2 years in the total population.Conclusion: Increasing STAN usage provided consistent improvements in fetal outcome equalling those noted in the Swedish randomized controlled trial (RCT) without increasing operative interventions for fetal distress. (c) 2006 Mosby, Inc. All rights reserved.