Multiple Moving Membrane CMUT With Enlarged Membrane Displacement and Low Pull-Down Voltage
Multiple Moving Membrane CMUT With Enlarged Membrane Displacement and Low Pull-Down Voltage
复制标题
DOI:
10.1109/led.2013.2286902
复制
发表时间:
2013-11
影响因子:
4.9
通讯作者:
T. Emadi;D. Buchanan
中科院分区:
文献类型:
--
作者:
T. Emadi;D. Buchanan
A multiple moving membrane capacitive micromachined ultrasonic transducer ( M3-CMUT) has been fabricated and is shown to exhibit a significantly enlarged total membrane displacement ( ~ 280 nm) compared with the displacement of a conventional CMUT ( ~ 85 nm) for the same bias voltage. The M3-CMUT exhibits a significant reduction of the device pull-down voltage, while showing a much higher capacitance change, 100 fF for the M3-CMUT compared with only 20 fF for CMUT, at a dc bias of 25 V. The device performance, sensitivity, and acoustic power generation capability are primarily associated with the magnitude of the displacement of the membrane, and therefore, are enhanced through employing multiple deflectable membranes in M3-CMUT device. This high performance M3-CMUT is a promising candidate for high resolution ultrasonic imaging application.