Type and accuracy of sphygmomanometers in primary care: a cross-sectional observational study
Type and accuracy of sphygmomanometers in primary care: a cross-sectional observational study
复制标题
DOI:
10.3399/bjgp11x593884
复制
发表时间:
2011-09-01
影响因子:
5.9
通讯作者:
Heneghan, Carl
中科院分区:
文献类型:
--
作者:
A'Court, Christine;Stevens, Richard;Heneghan, Carl
BackgroundPrevious studies identified worrying levels of sphygmomanometer inaccuracy and have not been repeated in the era of digital measurement of blood pressureAimTo establish the type and accuracy of sphygmomanometers in current useDesign and settingCross-sectional, observational study in 38 Oxfordshire primary care practicesMethodSphygmomanometers were evaluated between 50 and 250 mmHg, using Omron PA350 or Scandmed 950831-2 pressure meters.ResultsSix hundred and four sphygmomanometers were identified: 323 digital (53%), 192 aneroid (32%), 79 mercury (13%), and 10 hybrid (2%) devices. Of these, 584 (97%) could be fully tested. Overall, 503/584 (86%) were within 3 mmHg of the reference, 77/584 (13%) had one or more error of 4-9 mmHg, and 4/584 (