A sensitive and selective preconcentration method using solid-phase extraction (SPE) disk made from oxidized multiwalled carbon nanotubes (OMWCNTs), has been developed for the determination of aniline derivatives, such as 2-nitroaniline (2-NA), 4-nitroaniline (4-NA), and 2,4-dichloroaniline (2,4-DCA) in water samples. Anilines were extracted onto OMWCNT disk and then determined by high performance liquid chromatography (HPLC) with UV detector. Several parameters on the recovery of the analytes were investigated. The experimental results showed that it was possible to obtain quantitative analysis when the solution pH was 8 using 200 mL of validation solution containing 2 μg of anilines and 10 mL of acetonitrile/ethanol (8/2, v/v) as an eluent. Relative standard deviations for five determinations were 7.5% (2-NA), 6.5% (4-NA) and 3.8% (2,4-DCA) under optimum conditions. The linear range of calibration curves were 0.5 ng·mL−1to 15 ng·mL−1for each analyte with good correlation coefficients. The detection limits (3S/N) of 2-NA, 4-NA and 2,4-DCA were 30 pg·mL−1, 31 pg·mL−1and 26 pg·mL−1, respectively. Our method was successfully applied to the determination of aniline compounds in river water sample with high precision and accuracy.