Note: low temperature superconductor superconducting quantum interference device system with wide pickup coil for detecting small metallic particles.
Note: low temperature superconductor superconducting quantum interference device system with wide pickup coil for detecting small metallic particles.
复制标题
注:低温超导超导量子干涉装置系统具有宽拾波线圈,用于检测微小金属颗粒。
DOI:
10.1063/1.4739311
复制
发表时间:
2012
期刊:
影响因子:
--
通讯作者:
Y. Seki
中科院分区:
文献类型:
--
作者:
A. Kandori;K. Ogata;R. Kawabata;S. Tanimoto;Y. Seki
A one-channel low temperature superconductor superconducting quantum interference device system comprising a second-order axial gradiometer with a sensing area of 10 mm × 190 mm was developed. The gradiometer was mounted in a liquid-helium dewar (450-mm diameter; 975-mm length), with a gap of 12 mm between the pickup coil and the dewar-tail surface. The magnetic field sensitivity was measured to be 16 fT/Hz(1/2) in the white noise regime above 2 Hz. The system was used to measure stainless steel particles of different sizes passing through the sensing area. A 100-μm diameter SUS304 particle was readily detected passing at different positions underneath the large pickup coil by measuring its 1.3-pT magnetic field. Thus, the system was shown to be applicable to quality control of lamination sheet products such as lithium ion batteries.