High-resolution MRI vs multislice spiral CT: Which technique depicts the trabecular bone structure best?
High-resolution MRI vs multislice spiral CT: Which technique depicts the trabecular bone structure best?
复制标题
DOI:
10.1007/s00330-002-1695-5
复制
发表时间:
2003-04-01
影响因子:
5.9
通讯作者:
Majumdar, S
中科院分区:
文献类型:
--
作者:
Link, TM;Vieth, V;Majumdar, S
The purpose of this study was to compare trabecular bone structure parameters obtained from high-resolution magnetic resonance (HRMR) and multislice computed tomography (MSCT) images with those determined in contact radiographs from corresponding specimen sections. High-resolution MR and MSCT images were obtained in 39 distal radius specimens. For HRMR the in-plane spatial resolution was 0.152 x 0.153 mm(2) with a slice thickness of 0.9 and 0.3 mm using a 3D T1-weighted spin-echo sequence. For MSCT the resolution was 0.247 x 0.247 mm(2) with a collimation of 1 mm. Using a diamond saw, 117 0.9- to 1-mm-thick sections were obtained from these specimens and contact radiographs were acquired. In the corresponding sections structure parameters analogous to bone histomorphometry were determined. Significant correlations between MR- and CT-derived structure parameters and those derived from the contact radiographs were found (p