3-D FLASH IMAGING USING A SINGLE SURFACE COIL AND A NEW ADIABATIC PULSE, BIR-4
3-D FLASH IMAGING USING A SINGLE SURFACE COIL AND A NEW ADIABATIC PULSE, BIR-4
复制标题
DOI:
10.1097/00004424-199005000-00015
复制
发表时间:
1990-05-01
影响因子:
6.7
通讯作者:
GARWOOD, M
中科院分区:
文献类型:
--
作者:
STAEWEN, RS;JOHNSON, AJ;GARWOOD, M
A new adiabatic pulse, which can induce uniform and arbitrary flip angles despite the presnce of transmitter coil magnetic field (B1) inhomogeneities, is employed for 3-D fast imaging using a single surface coil for pulse transmission and signal detection. Computer calculations and phantom, rat, and human surface coil imaging experiments demonstrate the utility of this adiabatic pulse for T1-weighted imaging with a transmitter coil which generates a highly inhomogeneous B1 field profile.