ACCURACY OF PHASE-CONTRAST FLOW MEASUREMENTS IN THE PRESENCE OF PARTIAL-VOLUME EFFECTS
ACCURACY OF PHASE-CONTRAST FLOW MEASUREMENTS IN THE PRESENCE OF PARTIAL-VOLUME EFFECTS
复制标题
DOI:
10.1002/jmri.1880030213
复制
发表时间:
1993-03-01
期刊:
影响因子:
--
通讯作者:
PARKER, DL
中科院分区:
文献类型:
--
作者:
TANG, C;BLATTER, DD;PARKER, DL
The accuracy of volume flow rate measurements obtained with phase-contrast methods was assessed by means of computer simulation and in vitro experiments. Factors studied include (a) the partial-volume effect due to voxel dimensions relative to vessel dimensions and orientation and (b) intravoxel phase dispersion. It is shown that limited resolution (partial-volume effect) is the major obstacle to accurate flow measurement for both laminar and plug flow. The results show that at least 1 6 voxels must cover the cross section of the vessel lumen to obtain a measurement accuracy to within 10%. Measurement accuracy also greatly depends on the relative signal intensity of stationary tissue and is better for laminar flow than plug flow.