Novel separation mode of HPLC based on phase-separation multiphase flow
Novel separation mode of HPLC based on phase-separation multiphase flow
复制标题
DOI:
10.1007/s44211-022-00127-1
复制
发表时间:
2022-06-01
影响因子:
1.6
通讯作者:
Tsukagoshi, Kazuhiko
中科院分区:
文献类型:
--
作者:
Horikawa, Kosei;Kinoshita, Satoru;Tsukagoshi, Kazuhiko
A novel separation mode for high-performance liquid chromatography (HPLC) is proposed based on phase-separation multiphase flow. A commercially available HPLC system was used with a packed-separation column of octadecyl-silica (ODS)-modified particles. Water/acetonitrile/ethyl acetate ternary mixed solutions, (a) 1:8:1, (b) 1:3:1, and (c) 16:3:1 (v/v/v), were delivered into the system as an eluent at 20 and 5 degrees C. The ternary mixed solution acted as a two-phase separation solution leading to phase-separation multiphase flow. The solution flowed in the column homogeneously and heterogeneously at 20 and 5 degrees C, respectively. 1-Naphthol (NA) and 2,6-naphthalenedisulfonic acid (NDS) were injected into the system as model analytes. At 20 degrees C, the analyte mixture did not separate in solutions (a) and (b) while it separated in solution (c) with the elution order of NDS followed by NA. At 5 degrees C, it did not separate in solution (a), while it separated in solution (b) with elution order of NA followed by NDS and solution (c) with elution order of NDS followed by NA more effectively than 20 degrees C. The separation behavior and elution order are possibly caused by the phase-separation multiphase flow.