On the linearity of the MINOS light-injection calibration system
On the linearity of the MINOS light-injection calibration system
复制标题
MINOS光注入校准系统的线性度
DOI:
10.1016/j.nima.2003.10.105
复制
发表时间:
2004
影响因子:
1.4
通讯作者:
R. White
中科院分区:
文献类型:
--
作者:
P. Adamson;L. Barrett;A. Belias;A. Blake;A. Cabrera;E. Falk;P. Harris;J. Hartnell;C. Howcroft;M. Kordosky;K. Lang;R. Lee;J. Liu;D. Michael;R. Morse;J. Nelson;R. Nichol;G. Pearce;K. Ruddick;P. N. Smith;P. Symes;Julian Thomas;P. Vahle;R. Webb;A. Weber;R. White
The MINOS light-injection calibration system has been fully documented in a previous article (Nucl. Instr. and Meth. A 492 (2002) 353). Upon commissioning in the MINOS detectors, the system was found to give a non-linear response to variations in the intensity of injected light. The source of this non-linearity has been traced to a small change in the spectrum of the injected light as a function of the current applied to the original blue LEDs, in combination with a rapidly varying spectral response function of the wavelength-shifting fibre used in the detector. Both aspects of the problem have been addressed successfully by use of LEDs with different spectral characteristics, and the system now has a linear response.