Evaluation of the Magiscan Image Analyzer for Asbestos Fiber Counting
Evaluation of the Magiscan Image Analyzer for Asbestos Fiber Counting
复制标题
用于石棉纤维计数的 Magiscan 图像分析仪的评估
DOI:
10.1080/15298668791384346
复制
发表时间:
1987
期刊:
影响因子:
--
通讯作者:
S. Shulman
中科院分区:
文献类型:
--
作者:
P. Baron;S. Shulman
The Magiscan 2 (M-2) image analysis system with asbestos fiber counting software was evaluated. The M-2 takes a video image from a standard phase contrast light microscope used for human (manual) counting of asbestos fibers, processes the video image and then analyzes the particle shapes to count the number of fibers. Operator attention is required to the extent of placing the slide in the microscope, selecting the areas for analysis and focusing the microscope. A set of Occupational Safety and Health Administration (OSHA) quality control field samples was analyzed by the M-2, and the results were compared to the OSHA counts. The data indicated good precision by the M-2 and reasonably good agreement with the OSHA counts. A subset of these samples also was analyzed by a group of 12 proficient laboratories. The M-2 results agreed well with the inter-laboratory means. Finally, the performance of the M-2 was observed for individual fibers. The M-2 did not always agree with the observer (e.g., sometimes missin...