An evaluation in the diabetes clinic of the Cholestech L‐D‐X analyser
An evaluation in the diabetes clinic of the Cholestech L‐D‐X analyser
复制标题
Cholestech L-D-X 分析仪在糖尿病诊所的评估
DOI:
10.1002/pdi.1960110406
复制
发表时间:
1994
期刊:
影响因子:
--
通讯作者:
B. Slavin
中科院分区:
文献类型:
--
作者:
M. Cummings;G. Watts;J. Mazurkiewicz;B. Slavin
The Cholestech L-D-X lipid analyser (LDX) is one of the first fully automated desktop analysers to determine serum high-density-lipoprotein cholesterol (HDL-C) concentration, an important measurement for assessing dyslipidaemia, particularly in patients with non-insulin-dependent diabetes mellitus. We compared results of total cholesterol (TC), triglyceride (TG) and HDL-C concentrations measured by the LDX and by a standard (DAX) laboratory analyser from serum obtained from 40 patients in our diabetic clinic. The mean differences between the concentrations measured were not statistically significant, there being a very close correlation between methods (TC: LDX = 0.98. DAX + 0.06, r = 0.98, p<0.01; TG: LDX = 1.01. DAX + 0.05, r = 0.99, p<0.01 and HDL -C: LDX = DAX + 0.03, r = 0.97, p<0.01). Random error assessed by analysis of variance was minimal. We conclude that the Cholestech L-D-X analyser may be useful for providing rapid and accurate measurements of serum lipid profiles, particularly in patients in whom measuring HDL-C concentration is important. The small size, speed of analysis and simplicity of use of the LDX make it an ideal instrument for utilisation in the diabetes clinic and lipid out-patient clinic.