Comparison of improved precipitation methods for quantification of high-density lipoprotein cholesterol.
Comparison of improved precipitation methods for quantification of high-density lipoprotein cholesterol.
复制标题
DOI:
10.1093/clinchem/31.2.217
复制
发表时间:
1985-02
影响因子:
9.3
通讯作者:
Russell Warnicl;Thuy Nguyen;Alegrla A. Albers
中科院分区:
文献类型:
--
作者:
Russell Warnicl;Thuy Nguyen;Alegrla A. Albers
We compared the standard Lipid Research Clinics heparin-Mn2+ (46 mmol/L) method and five improved precipitation methods for quantification of high-density lipoprotein (HDL) cholesterol. Three of these methods--a dextran sulfate-Mg2+ procedure, reported as a Selected Method, a modified heparin-Mn2+ (92 mmol/L) method, and a modified phosphotungstate-Mg2+ procedure--all gave similar results. Three other methods--the standard heparin-Mn2+ (46 mmol/L) method and two polyethylene glycol methods (75 g/L or pH 10 reagent at 100 g/L final concentrations)--gave slightly higher values for HDL cholesterol. Addition of NaCl or glucose to specimens did not significantly change protein precipitation. In terms of sedimentation effectiveness with hypertriglyceridemic specimens, the methods were ranked in the following order: polyethylene glycol (pH 10, 100 g/L) greater than dextran sulfate-Mg2+ greater than heparin-Mn2+ (92 mmol/L) = polyethylene glycol (75 g/L) greater than phosphotungstate-Mg2+ greater than heparin-Mn2+ (46 mmol/L).