Final report on CIPM key comparison CCAUV.V-K2
Final report on CIPM key comparison CCAUV.V-K2
复制标题
CIPM 关键比较 CCAUV.V-K2 最终报告
DOI:
10.1088/0026-1394/51/1a/09002
复制
发表时间:
2014
期刊:
影响因子:
2.4
通讯作者:
A. Täubner
中科院分区:
文献类型:
--
作者:
T. Bruns;G. Ripper;A. Täubner
The results of the third CIPM comparison in the area of vibration, here concerning sinusoidal acceleration, are presented. This comparison was intended to build the foundation for subsequent regional key comparisons over an extended frequency range. However, one of the transducers showed a poor stability. This feature, and other shortcomings that became apparent during the analysis, limit the impact of the comparison results. Therefore, a key comparison reference value has been defined for a limited domain and application, and the respective Degrees of Equivalence are reported. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).