在卢森堡Walferdange地下地球动力学实验室(WULG)对A10-008绝对重力仪的设定标准偏差、可重复性和偏移量进行了评估。数据分析表明,该仪器在制造商的规格范围内运行。对于A10-008,平均设定标准差为(1.6±0.6)µGal (1Gal≡1 cm s−2),平均重复性为(2.9±1.5)µGal,与绝对重力仪FG5-216相比的平均偏移量为(3.2±3.5)µGal。
The set standard deviation, repeatability and offset of absolute gravimeter A10-008 were assessed at the Walferdange Underground Laboratory for Geodynamics (WULG) in Luxembourg. Analysis of the data indicates that the instrument performed within the specifications of the manufacturer. For A10-008, the average set standard deviation was (1.6 ± 0.6) µGal (1Gal ≡ 1 cm s−2), the average repeatability was (2.9 ± 1.5) µGal, and the average offset compared to absolute gravimeter FG5-216 was (3.2 ± 3.5) µGal.