MEASUREMENT OF PHENYLALANINE IN ROUTINE CARE OF PHENYLKETONURIC CHILDREN
MEASUREMENT OF PHENYLALANINE IN ROUTINE CARE OF PHENYLKETONURIC CHILDREN
复制标题
DOI:
10.1136/adc.48.6.472
复制
发表时间:
1973-01-01
影响因子:
5.2
通讯作者:
FARQUHAR, JW
中科院分区:
文献类型:
--
作者:
BELTON, NR;CROMBIE, JD;FARQUHAR, JW
Fluorimetry, paper and column chromatography, and bacterial inhibition assay have been compared in the quantitation of serum or blood phenylalanine at levels ranging from 1 to over 30 mg/100 ml. Special attention was paid to levels of 2 mg or less and to those in excess of 20 mg/100 ml, since these have therapeutic implications. Bacterial inhibition assay, under routine conditions, tended to read low or not at all at 2 mg or less, and to read high above 20 mg/100 ml so that it `failed safe' under such conditions. Paper chromatography was shown to be a simple, rapid, and economic method of getting accurate results at all levels in the range examined.