X-ray polarimetry with a conventional gas proportional counter through rise-time analysis
X-ray polarimetry with a conventional gas proportional counter through rise-time analysis
复制标题
使用传统气体正比计数器通过上升时间分析进行 X 射线偏振测定
DOI:
10.1016/s0168-9002(98)01145-0
复制
发表时间:
1999
影响因子:
1.4
通讯作者:
K. Tamura
中科院分区:
文献类型:
--
作者:
K. Hayashida;N. Miura;H. Tsunemi;K. Torii;H. Murakami;Y. Ohno;K. Tamura
We have performed an experiment on the signal rise time of a Xe gas proportional counter using a polarized X-ray beam of synchrotron orbital radiation with energies from 10 to 40keV. When the counter anode is perpendicular to the electric vector of the incident X-ray photons, the average rise time becomes significantly longer than that for the parallel case. This indicates that the conventional gas proportional counters are useful for X-ray polarimetry. The moderate modulation contrast of this rise-time polarimeter (M=0.1 for 10keV X-rays and M=0.35 for 40keV X-rays), with capability of the simultaneous measuring X-ray energies and the timing, would be useful for applications in X-ray astronomy and in other fields.