A comparison of techniques for size measurement of nanoparticles in cell culture medium

A comparison of techniques for size measurement of nanoparticles in cell culture medium
复制标题

DOI:
10.1039/c6ay00419a
复制
发表时间:
2016-01-01
期刊:
影响因子:
3.1
通讯作者:
Varga, Zoltan
Varga, Zoltan
中科院分区:
化学3区
文献类型:
--
作者:
Gollwitzer, Christian;Bartczak, Dorota;Varga, Zoltan

文献摘要

被引文献

相似文献

Plain and aminated silica nanoparticles dispersed in purified water, in 50 mM Tris-HCl buffer and in cell culture medium were measured using dynamic light scattering (DLS), centrifugal liquid sedimentation (CLS), small-angle X-ray scattering (SAXS), and particle tracking analysis (PTA). The test samples were measured by all methods immediately after dispersion and after incubation at room temperature for 24 h. The effect of the biological dispersion medium on the modal value of the particle size distribution was compared for each method taking into account the estimated uncertainty. For the methods based on light scattering, DLS and PTA, the size distributions obtained were significantly altered due to the formation of a protein corona and induced agglomeration effects. With SAXS and CLS, the measured size of the primary particles was mostly unchanged. While SAXS offers excellent precision and traceability to the SI unit system if the model fitting approach is used for data analysis, CLS provides detailed size distributions from which additional information on the agglomeration state can be deduced.