A novel autonomic activation measurement method for stress monitoring: non-contact measurement of heart rate variability using a compact microwave radar
A novel autonomic activation measurement method for stress monitoring: non-contact measurement of heart rate variability using a compact microwave radar
复制标题
DOI:
10.1007/s11517-007-0298-3
复制
发表时间:
2008-07-01
影响因子:
3.2
通讯作者:
Kawakami, Mitsuyuki
中科院分区:
文献类型:
--
作者:
Suzuki, Satoshi;Matsui, Takemi;Kawakami, Mitsuyuki
We developed a novel method for non-contact monitoring of stress-induced autonomic activation through the back of a chair, using a compact 24 GHz microwave radar (8 x 5 x 3 cm), without large-scale equipment and placing a heavy burden on the monitored individual. Following a silent period of 120 s, audio stimuli using a composite tone of 2,120 and 2,130 Hz sine-waves at 95 dB were conducted for 120 s. From dorsal, LF/HF of HRV reflecting sympatho-vagal balance was determined by microwave radar with the maximum entropy method using eight volunteers (mean age 23 +/- 1 years). Mean LF/HF measured by non-contact and contact (using electrocardiography for reference) methods during audio stimuli increased 34 and 37%, respectively, as compared with those of the silent period. Maximum cross-correlations between contact and non-contact measurements averaged 0.73 +/- 0.10. Our method appears to be promising for future monitoring of stress-induced autonomic activation of operators and may reduce stress-induced accidents.