Film dosimetry of megavoltage photon beams: a practical method of isodensity-to-isodose curve conversion.
Film dosimetry of megavoltage photon beams: a practical method of isodensity-to-isodose curve conversion.
复制标题
兆伏光子束的薄膜剂量测定:等密度到等剂量曲线转换的实用方法。
DOI:
10.1118/1.594913
复制
发表时间:
1981
期刊:
影响因子:
3.8
通讯作者:
Sharma,SC
中科院分区:
文献类型:
--
作者:
Williamson,JF;Khan,FM;Sharma,SC
The central problems of photon beam film dosimetry are the dependence of film response upon photon energy, processing conditions, and film plane orientation. We have overcome these problems by accurately fitting the depth‐dependent sensitometric curve of Kodak XV‐2 film (exposed parallel to beam axis) to the equation OD(D,d) =ODs(1 − exp {−α0[1+β(d−dm)]D}) where OD(D,d) is the optical density for doseDat depthd. ODs, α0, and β are constants characteristic of the film and beam energy but are independent of field size. Only central axis depth dose data for a single field are required to determine their values. A computer program based upon this equation has been written which successfully generates single field isodose curves from film data for a variety of field sizes (including wedged fields) with an accuracy of ±3%. Data are presented for60Co, and 4 and 10 MV x rays.