Possibility of real-time measurement of an airborne Cryptomeria japonica pollen allergen based on the principle of surface plasmon resonance
Possibility of real-time measurement of an airborne Cryptomeria japonica pollen allergen based on the principle of surface plasmon resonance
复制标题
基于表面等离子体共振原理实时测量空气中柳杉花粉过敏原的可能性
DOI:
10.1023/a:1013002001583
复制
发表时间:
2001
期刊:
影响因子:
2
通讯作者:
H. Nitta
中科院分区:
文献类型:
--
作者:
Yuichi Takahashi;T. Ohashi;T. Nagoya;M. Sakaguchi;H. Yasueda;H. Nitta
In order to establish a real-timemeasurement system for the concentration ofairborne pollen allergens, we examined Cry j 1,one of the major allergens ofCryptomeriajaponicapollen, as an example to establishthe system. The feasible system consisted of:collection of airborne pollen allergens usingthe Virtual Impactor or Cyclone sampler,extraction of Cry j 1 using 10mM HEPES buffercontaining 0.125M NH4HCO3andfinally, real-time measurement of Cry j 1 usingthe BIACORE 3000 system. The sensitivity of thesystem was 5 ng/ml, and 0.1 ml sample volumeand at least 500 pg of Cry j 1 were requiredfor each measurement. Quantification of Cry j 1in the air can be determined 30 min aftercollection, i.e. 15 min for extraction, 10 minfor separation from particulate matters and 5min for the measurement.