Effective Dose during Abdominal Three-dimensional Imaging with a Flat-Panel Detector Angiography System
Effective Dose during Abdominal Three-dimensional Imaging with a Flat-Panel Detector Angiography System
复制标题
DOI:
10.1148/radiol.2502080695
复制
发表时间:
2009-02-01
期刊:
影响因子:
19.7
通讯作者:
Kobayashi, Ikuo
中科院分区:
文献类型:
--
作者:
Suzuki, Shigeru;Furui, Shigeru;Kobayashi, Ikuo
The purpose of this study was to measure the effective dose during abdominal three-dimensional imaging obtained with an angiography unit with a digital flat-panel system on a phantom and to determine dose-area product (DAP)-to-effective dose conversion factors. DAPs and effective doses were evaluated for 163-cm-tall human-shaped phantoms with estimated body weights of 54, 64, and 77 kg, and the effective doses were 2.1, 3.2, and 4.2 mSv, respectively. The DAP-to-effective dose conversion factors were 0.28-0.29 mSv . Gy(-1) . cm(-2). In conclusion, the DAPs were useful for estimating the effective dose during abdominal three-dimensional angiographic imaging. (C) RSNA, 2009