Fast voice-coil scanning optical-resolution photoacoustic microscopy.
Fast voice-coil scanning optical-resolution photoacoustic microscopy.
复制标题
DOI:
10.1364/ol.36.000139
复制
发表时间:
2011-01-15
期刊:
影响因子:
3.6
通讯作者:
Wang LV
中科院分区:
文献类型:
--
作者:
Wang L;Maslov K;Yao J;Rao B;Wang LV
We developed a photoacoustic imaging system that has real-time imaging capability with optical resolution. The imaging system is capable of scanning at 20 Hz over a 9-mm range, and up to 40 Hz over a 1-mm scanning range. A focused laser beam provides a lateral resolution of 3.4 μm as measured in an optically nonscattering medium. Flows of micron-sized carbon particles or whole blood in a silicone tube and individual red blood cells (RBCs) in mouse ear capillaries were also imaged in real time, which demonstrated the capability to image highly dynamic processes in vivo at a micron-scale resolution.