A SU-8-Based Fully Integrated Biocompatible Inductively Powered Wireless Neurostimulator

A SU-8-Based Fully Integrated Biocompatible Inductively Powered Wireless Neurostimulator
复制标题

DOI:
10.1109/jmems.2012.2221155
复制
发表时间:
2013-02-01
影响因子:
2.7
通讯作者:
Lee, Jeong-Bong
Lee, Jeong-Bong
中科院分区:
工程技术3区
文献类型:
--
作者:
Cho, Sung-Hoon;Xue, Ning;Lee, Jeong-Bong

文献摘要

被引文献

相似文献

设计、制造了一种完全集成的微创紧凑型聚合物无线神经刺激器,并在露天和体内进行了表征。神经刺激器(3.1 x 1.6 x 0.3 mm)由一个平面螺旋线圈(用于通过电感耦合进行无线供电)、两个肖特基二极管(用于全波整流)、一个专用集成电路神经刺激器电路芯片(用于产生刺激尖峰信号)和两个双相铂铱(PtIr)刺激电极组成。该器械完全集成并完全嵌入生物相容性SU-8包装中。对于体内测试,将无线神经刺激器皮下植入大鼠后肢。在394-MHz谐振频率下,在21 dBm(125 mW)的耦合功率下,记录了在延长的无线刺激期间稳定和稳健的皮质反应。
A fully integrated minimally invasive compact polymer-based wireless neurostimulator was designed, fabricated, and characterized both in open air and in vivo. The neurostimulator (3.1 x 1.6 x 0.3 mm) consists of a planar spiral coil for wireless power supply through inductive coupling, two Schottky diodes for full-wave rectification, an application-specific integrated circuit neurostimulator circuit chip for stimulus spike signal generation, and two biphasic platinum-iridium (PtIr) stimulation electrodes. The device is fully integrated and completely embedded in biocompatible SU-8 packaging. For in vivo testing, the wireless neurostimulator was implanted subcutaneously in a rat hind limb. At the coupling power of 21 dBm (125 mW) at 394-MHz resonant frequency, stable and robust cortical responses during extended periods of wireless stimulation were recorded.