Average lens density quantification with swept-source optical coherence tomography: optimized, automated cataract grading technique
Average lens density quantification with swept-source optical coherence tomography: optimized, automated cataract grading technique
复制标题
DOI:
10.1016/j.jcrs.2019.07.033
复制
发表时间:
2019-12-01
影响因子:
2.8
通讯作者:
Gatinel, Damien
中科院分区:
文献类型:
--
作者:
Panthier, Christophe;de Wazieres, Alice;Gatinel, Damien
Purpose: To assess a revised version of an objective cataract grading method based on the lens densitometry on swept source optical coherence tomography (SS-OCT) scans.Setting: Rothschild Foundation, Paris, France.Design: Prospective case series.Methods: All patients seeking a cataract evaluation and consenting to study participation were included. Exclusion criteria were a history of ocular surgery, cornea or retina disorders, and ocular dryness. The average lens densitometry (ALD) was measured on SS-OCT scans (IOLMaster 700) using an algorithm developed with MATLAB software. The ocular scatter index (OSI), measured using a double-pass aberrometer (Optical Quality Analysis System); Pentacam nucleus staging (PNS) score; and mean PNS were also measured and compared with the ALD measurements.Results: Two hundred eighty-five eyes (195 with cataract; 90 controls) of 155 patients aged 20 to 93 years (mean 56.5 years 21 [SD]) were included. The ALD measurements were repeatable (P = .99, analysis of variance). The repeatability limit was 2.50 pixel units. The ALD correlated with the OSI (r(2) = 0.41, P < .01) and PNS score (r(2) = 0.65, P