Development of Non-Contact Electrical Resistivity Measurement Technique Using an Electrostatic Levitator
Development of Non-Contact Electrical Resistivity Measurement Technique Using an Electrostatic Levitator
复制标题
DOI:
--
复制
发表时间:
2008-07
期刊:
影响因子:
--
通讯作者:
T. Ishikawa;P. Paradis;Y. Watanabe;N. Koike
中科院分区:
文献类型:
--
作者:
T. Ishikawa;P. Paradis;Y. Watanabe;N. Koike
A non-contact technique to measure the electrical resistivity of levitated samples is reported. The technique utilizes the principle of the asynchronous induction motor, measuring the induced torque by applying a rotating magnetic field to the sample. Relation between sample size and induced torque was experimentally checked with aluminum samples and proved that induced torque is proportional to the fifth power of the sample radius. Based on this result, the electrical resistivity of solid zirconium at high temperatures was measured and showed a good agreement with literature data.