The use of bushings with triple point of water cells: towards breaking the 50 µK barrier
The use of bushings with triple point of water cells: towards breaking the 50 µK barrier
复制标题
使用带有水三相点池的套管:突破 50 µK 屏障
DOI:
--
复制
发表时间:
2008
期刊:
影响因子:
--
通讯作者:
R. Dematteis
中科院分区:
文献类型:
--
作者:
P. Steur;R. Dematteis
Instabilities of the order of 50 µK or higher have been observed in water triple point cells with a thermometer well of 12 mm to 13.6 mm diameter during measurements at the INRiM in the framework of the Euromet.T-K7 comparison. In the course of many follow-up measurements, extending over more than one year, these anomalies have been investigated and a solution has been tried and verified. The use of quartz bushings has been tried, leading to a much closer agreement between water triple point cells of different dimensions than was hitherto achieved. Furthermore, the effect on the realized triple point temperature is shown, when no bushing is used.