LENS EQUIVALENT DOSE OF STAFF DURING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY: DOSE COMPARISON USING TWO TYPES OF DOSEMETERS
LENS EQUIVALENT DOSE OF STAFF DURING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY: DOSE COMPARISON USING TWO TYPES OF DOSEMETERS
复制标题
内镜逆行胰胆管造影期间工作人员的镜片等效剂量:使用两种类型剂量计的剂量比较
DOI:
10.1093/rpd/ncac160
复制
发表时间:
2022
影响因子:
1
通讯作者:
Hiroo Chiba and Hiroshi Fukuda
中科院分区:
文献类型:
--
作者:
Yoshiaki Morishima;Koichi Chida; Takayoshi Meguro;Morihisa Hirota;Hiroo Chiba and Hiroshi Fukuda
This study aimed to compare the lens equivalent dose (LED) measured during endoscopic retrograde cholangiopancreatography (ERCP) using DOSIRIS™ as a dedicated dosemeter to that measured using glass badges to determine if glass badges can be alternative tools for LED measurement.LEDs for physicians during ERCP were measured using the DOSIRIS™ [3-mm dose equivalent] worn on the outer edge of the eyes and personal dosemeters (glass badges) [0.07-mm dose equivalent] worn on the right and left sides of the neck.The cumulated doses over 6 months for the left eye using DOSIRIS™ were 9.5 and 11.8 mSv for physicians A and B, whereas doses measured using glass badges were 7.5 and 11.6 mSv, respectively. The LEDs of the physicians at the left eye and left neck side showed almost similar values and were significantly correlated (r = 0.95;p< 0.01).For an accurate LED measurement during ERCP, using a dosemeter such as DOSIRIS™ is recommended, although similar LED estimation values were reported using glass badges on the left neck side.