Fast Fourier transform spectrometer readout for large arrays of microwave kinetic inductance detectors
Fast Fourier transform spectrometer readout for large arrays of microwave kinetic inductance detectors
复制标题
用于大型微波动能电感探测器阵列的快速傅立叶变换光谱仪读数
DOI:
--
复制
发表时间:
2009
期刊:
影响因子:
--
通讯作者:
R. Güsten
中科院分区:
文献类型:
--
作者:
S. Yates;A. Baryshev;J. Baselmans;B. Klein;R. Güsten
Microwave kinetic inductance detectors have great potential for large, very sensitive detector arrays for use in, for example, submillimeter imaging. Being intrinsically readout in the frequency domain, they are particularly suited for frequency domain multiplexing allowing ∼1000 s of devices to be readout with one pair of coaxial cables. However, this moves the complexity of the detector from the cryogenics to the warm electronics. We present here the concept and experimental demonstration of the use of fast Fourier transform spectrometer readout, showing no deterioration of the noise performance compared to the low noise analog mixing while allowing high multiplexing ratios.