Mammographic imaging with a small format CCD-based digital cassette: Physical characteristics of a clinical system
Mammographic imaging with a small format CCD-based digital cassette: Physical characteristics of a clinical system
复制标题
DOI:
10.1118/1.1286720
复制
发表时间:
2000-08-01
期刊:
影响因子:
3.8
通讯作者:
D'Orsi, CJ
中科院分区:
文献类型:
--
作者:
Vedantham, S;Karellas, A;D'Orsi, CJ
The physical characteristics of a clinical charge coupled device (CCD)-based imager (Senovision, GE Medical Systems, Milwaukee, WI) for small-field digital mammography have been investigated. The imager employs a MinR 2000(TM) (Eastman Kodak Company, Rochester, NY) scintillator coupled by a 1:1 optical fiber to a front-illuminated 61X61 mm CCD operating at a pixel pitch of 30 microns. Objective criteria such as modulation transfer function (MTF), noise power spectrum (NPS), detective quantum efficiency (DQE), and noise equivalent quanta (NEQ) were employed for this evaluation. The results demonstrated a limiting spatial resolution (10% MTF) of 10 cy/mm. The measured DQE of the current prototype utilizing a 28 kVp, Mo-Mo spectrum beam hardened with 4.5 cm Lucite is similar to 40% at close to zero spatial frequency at an exposure of 8.2 mR, and decreases to similar to 28% at a low exposure of 1.1 mR. Detector element nonuniformity and electronic gain variations were not significant after appropriate calibration and software corrections. The response of the imager was linear and did not exhibit signal saturation under tested exposure conditions. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)01308-0].