A device for dried blood microsampling in quantitative bioanalysis: overcoming the issues associated with blood hematocrit
A device for dried blood microsampling in quantitative bioanalysis: overcoming the issues associated with blood hematocrit
复制标题
DOI:
10.4155/bio.14.310
复制
发表时间:
2015-01-01
期刊:
影响因子:
1.8
通讯作者:
Rudge, James B.
中科院分区:
文献类型:
--
作者:
Spooner, Neil;Denniff, Philip;Rudge, James B.
Aims: A cross-laboratory experiment has been performed on a novel dried blood sampler in order to investigate whether it overcomes issues associated with blood volume and hematocrit (HCT) that are observed when taking a subpunch from dried blood spot samples. Materials & methods: An average blood volume of 10.6 l was absorbed by the samplers across the different HCTs investigated (20-65%). Results: No notable change of volume absorbed was noted across the HCT range. Furthermore, the variation in blood sample volumes across six different laboratories was within acceptable limits. Conclusion: The novel volumetric absorptive microsampling device has the potential to deliver the advantages of dried blood spot sampling while overcoming some of the issues associated with the technology.