On the relative efficacy of electropermeation and isothermal desorption approaches for measuring hydrogen diffusivity
On the relative efficacy of electropermeation and isothermal desorption approaches for measuring hydrogen diffusivity
复制标题
DOI:
10.1016/j.ijhydene.2022.10.025
复制
发表时间:
2022-10
影响因子:
7.2
通讯作者:
A. Zafra;Z. Harris;E. Korec;E. Mart'inez-Paneda
中科院分区:
文献类型:
--
作者:
A. Zafra;Z. Harris;E. Korec;E. Mart'inez-Paneda
The relative efficacy of electrochemical permeation (EP) and isothermal desorption spectroscopy (ITDS) methods for determining the hydrogen diffusivity is investigated using cold-rolled pure iron. The diffusivities determined from 13 first transient and 8 second transient EP experiments, evaluated using the conventional lag and breakthrough time methods, are compared to the results of 10 ITDS experiments. Results demonstrate that the average diffusivity is similar between the second EP transient and ITDS, which are distinctly increased relative to the first EP transient. However, the coefficient of variation for the ITDS experiments is reduced by 2 and 3-fold relative to the first and second EP transients, confirming the improved repeatability of ITDS diffusivity measurements. The source of the increased error in EP measurements is systematically evaluated, revealing an important influence of assumed electrochemical boundary conditions on the analysis and interpretation of EP experiments.