Construction of a dilution refrigerator based ultra-low temperature scanning tunneling microscope
Construction of a dilution refrigerator based ultra-low temperature scanning tunneling microscope
复制标题
DOI:
10.1143/jjap.45.1909
复制
发表时间:
2006-03-01
期刊:
影响因子:
--
通讯作者:
Fukuyama, H
中科院分区:
文献类型:
--
作者:
Kambara, H;Matsui, T;Fukuyama, H
We constructed a dilution refrigerator based ultra-low temperature scanning tunneling microscope (ULT-STM) which works at temperatures down to 20mK, in magnetic fields LIP to 6T and ill Ultrahigh vacuum (UHV). One call load samples/tips, which are prepared in a UHV chamber, to all STM head maintaining the low temperature and UHV conditions. After then they call be cooled back to the base temperature in several hours. We report results of a test measurement on a Superconducting NbSe2 sample as well as recent STM/STS studies oil graphite samples Such as observations of the Landau quantization and visualization of the possible localized states in magnetic fields.