Barker coded excitation with linear frequency modulated carrier for ultrasonic imaging
Barker coded excitation with linear frequency modulated carrier for ultrasonic imaging
复制标题
用于超声成像的线性调频载波巴克编码激励
DOI:
10.1016/j.bspc.2014.06.004
复制
发表时间:
2014-09-01
影响因子:
5.1
通讯作者:
Feng, Yizhi
中科院分区:
文献类型:
--
作者:
Fu, Juan;Wei, Gang;Feng, Yizhi
A new Barker coded excitation using linear frequency modulated (LFM) carrier (called LFM-Barker) is proposed for improving ultrasound imaging quality in terms of axial resolution and signal-to-noise ratio (SNR). The LFM-Barker coded excitation has two independent parameters: one is the bandwidth of LFM carrier, and the other is the chip duration of Barker code. To improve the axial resolution, increase the bandwidth of LEM carrier; and to improve the SNR, increase the chip duration of Barker code. In this study, a LFM pulse with proper (