The wavelength dependence of the response of a pulse fluorometer using the single photoelectron counting method
The wavelength dependence of the response of a pulse fluorometer using the single photoelectron counting method
复制标题
使用单光电子计数方法的脉冲荧光计响应的波长依赖性
DOI:
10.1063/1.1686442
复制
发表时间:
1974
影响因子:
1.6
通讯作者:
B. Donzel
中科院分区:
文献类型:
--
作者:
P. Wahl;J. Auchet;B. Donzel
We study the apparatus response function g (t) which must be known in order to deconvolute the fluorescence decay experiments obtained by the single photon counting method. Due to the flash and the PM characteristics, g (t) depends on the excitation and emission wavelengths. The PM contribution is analyzed qualitatively and found to be in agreement with the properties of the photoelectric effect. Therefore, the accurate determination of the ``true fluorescence decay'' cannot be obtained by the process commonly used, which consists in replacing g (t) by the apparatus response obtained with the PM receiving the light of the exciting pulse. The function g (t) may be obtained in using a reference fluorescent solution with a known fluorescence decay‐time.